Acoustic micro-tapping optical coherence elastography to quantify corneal collagen cross linking: ex vivo human study

Open AccessPublished:November 12, 2022DOI:https://doi.org/10.1016/j.xops.2022.100257

      Abstract

      Purpose

      To evaluate changes in the anisotropic elastic properties of ex vivo human cornea treated with UV cross-linking (CXL) using non-contact Acoustic micro-Tapping Optical Coherence Elastography (AμT-OCE).

      Design

      AμT-OCE was performed on normal and cross-linked human donor cornea in an ex vivo laboratory study.

      Subjects

      Normal human donor cornea (n= 22) divided into 4 subgroups. All samples were stored in optisol.

      Methods

      Elastic properties (in-plane Young’s, E, and out-of-plane, G, shear modulus) of normal and UV-CXL treated human corneas were quantified using non-contact AμT-OCE. A nearly incompressible transverse isotropic (NITI) model was used to reconstruct moduli from AμT-OCE data. Independently, cornea elastic moduli were also measured with destructive mechanical tests (tensile extensometry and shear rheometry).

      Main Outcome Measures

      Corneal elastic moduli (in-plane Young’s modulus, E, in-plane, μ, and out-of-plane, G, shear moduli) can be evaluated in both normal and CXL treated tissues, as well as monitored during the CXL procedure using non-contact AμT-OCE.

      Results

      CXL induced a significant increase in both in-plane and out-of-plane elastic moduli in human cornea. The statistical mean in the paired study (pre- and post-, n=7) of the in-plane Young’s modulus, E=3μ, increased from 19 MPa to 43 MPa while the out-of-plane shear modulus, G, increased from 188 kPa to 673 kPa. Mechanical tests in a separate subgroup support CXL-induced cornea moduli changes and generally agree with non-contact AμT-OCE measurements.

      Conclusions

      The human cornea is a highly anisotropic material where in-plane mechanical properties are very different from those out-of-plane. Non-contact AμT-OCE can measure changes in the anisotropic elastic properties in human cornea as a result of UV-CXL.

      Keywords

      The unique structure and organization of the collagen matrix within the cornea helps preserve shape and maintain transparency. Most corneal collagen is Type I, containing fibrils within the stroma that run the entire plane and connect with fibrils in the limbus or sclera.
      • Meek K.M.
      • Boote C.
      The organization of collagen in the corneal stroma.
      Collagen fibrils are more tightly packed in the anterior portion compared to the posterior, but are generally arranged within the stroma in 200 - 300 stacked sheets (1-2 μm thick) referred to as lamellae.
      • Koudouna E.
      • et al.
      Evolution of the vertebrate corneal stroma.
      The fibrils lie within a protein rich, hydrated proteoglycan mesh.
      • Borcherding M.S.
      • et al.
      Proteoglycans and collagen fibre organization in human corneoscleral tissue.
      ,
      • Scott J.E.
      • Thomlinson A.M.
      The structure of interfibrillar proteoglycan bridges ('shape modules’) in extracellular matrix of fibrous connective tissues and their stability in various chemical environments.
      It is well established that the complex micro-network of collagen bundles contributes to macro-scale deformation and stability linked to functional vision. In pathologies affecting the cornea, microstructural changes in the collagen fibers and proteoglycan-rich fluid of the stroma alter the cornea’s macro-scale biomechanical response to intraocular pressure
      • Meek K.M.
      • Knupp C.
      Corneal structure and transparency.
      and can result in corneal shape changes and degraded visual acuity.
      • Gandhi S.
      • Jain S.
      The anatomy and physiology of cornea.
      • Kotecha A.
      What Biomechanical Properties of the Cornea Are Relevant for the Clinician.
      • Dupps W.J.
      • Wilson S.E.
      Biomechanics and wound healing in the cornea.
      In degenerative corneal diseases such as keratoconus, local changes in mechanical properties result in a bulging conical shape associated with vision loss.
      • Kotecha A.
      What Biomechanical Properties of the Cornea Are Relevant for the Clinician.
      ,
      • Andreassen T.T.
      • Hjorth Simonsen A.
      • Oxlund H.
      Biomechanical Properties of Keratoconus and Normal Corneas.
      Keratoconus is commonly treated with collagen cross-linking (CXL), a procedure reported to delay ectasia progression by increasing stiffness via ultraviolet (UV) light induced crosslinks in stromal collagen soaked in riboflavin. While cross-linking can halt disease progression by greatly reducing the number of required keratoplasties,
      • Chan E.
      • Snibson G.R.
      Current status of corneal collagen cross-linking for keratoconus: A review.
      the specific relationships between enhanced cross-linking and mechanical stabilization in UV-CXL remain speculative,
      • Santhiago M.R.
      • Randleman J.B.
      The biology of corneal cross-linking derived from ultraviolet light and riboflavin.
      and both long-term and immediate outcomes remain unpredictable for individual patients.
      Due to these uncertainties, a personalized biomechanical model predicting post-surgical corneal shape and function is needed to improve CXL outcomes. Personalized biomechanical models of the cornea can be generated if (i) cornea shape and thickness are mapped and (ii) corneal elastic moduli are known.
      There are numerous clinical tools available to assess corneal shape, yet there are currently none that can accurately measure low-strain corneal elastic moduli pre-, intra- and post-surgery in a non-destructive, non-invasive manner. While technologies based on tonometry (e.g. Ocular Response Analyzer (ORA), Dynamic Scheimpflug Analyzer (DSA)) have shown some success in correlating biomechanical estimates with disease progression,
      • De Moraes C.
      Risk factors for visual field progression in treated glaucoma.
      ,
      • Roberts C.
      Concepts and misconceptions in corneal biomechanics.
      these methods rely on a number of novel metrics that indirectly assess elastic moduli responsible for corneal deformation.
      • Eliasy A.
      • et al.
      Determination of Corneal Biomechanical Behavior in-vivo for Healthy Eyes Using CorVis ST Tonometry: Stress-Strain Index.
      • Chong J.
      • Dupps W.J.
      Corneal biomechanics: Measurement and structural correlations.
      • Akram Abdelazim
      • Joda
      • et al.
      Development and validation of a correction equation for Corvis tonometry.
      • Hashemi H.
      • et al.
      Long-term Results of an Accelerated Corneal Cross-linking Protocol (18 mW/cm2) for the Treatment of Progressive Keratoconus.
      • Roberts Cynthia J.
      • et al.
      Introduction of two novel stiffness parameters and interpretation of air Puff-Induced biomechanical deformation parameters with a dynamic scheimpflug analyzer.
      Specifically, both use a relatively high-pressure, widely distributed air-puff to deform the cornea and measure its response. While ORA has been used in multiple CXL studies in vivo, the one-dimensional nature of the system has resulted in poor sensitivity in detecting differences between CXL and control groups, where several studies described stabilized disease progression but found no change in the novel biomechanics-related metrics of corneal hysteresis (CH) or corneal resistance factor (CRF).
      • Goldich Y.
      • et al.
      Clinical and corneal biomechanical changes after collagen cross-linking with riboflavin and uv irradiation in patients with progressive keratoconus: Results after 2 years of follow-up.
      ,
      • Vinciguerra P.
      • et al.
      Intra- and postoperative variation in ocular response analyzer parameters in keratoconic eyes after corneal cross-linking.
      Additionally, tonometry-based techniques often require a non-trivial IOP correction in finite element simulations
      • Eliasy A.
      • et al.
      Determination of Corneal Biomechanical Behavior in-vivo for Healthy Eyes Using CorVis ST Tonometry: Stress-Strain Index.
      ,
      • Elsheikh A.
      • et al.
      Stress free configuration of the human eye.
      and use a simplified isotropic mechanical model, leading to high variability depending on experimental conditions.
      Recently, optical coherence elastography (OCE) has measured corneal stiffness at various length scales while simultaneously mapping shape in a single non-invasive measurement.
      • Kirby M.A.
      • et al.
      Optical coherence elastography in ophthalmology.
      Contact-based OCE techniques have demonstrated potential to detect biomechanical differences between keratoconic and healthy corneas in vivo.
      • De Stefano V.S.
      • Ford M.R.
      • Seven I.
      • Dupps W.J.
      Depth-dependent corneal biomechanical properties in normal and keratoconic subjects by optical coherence elastography.
      However, direct contact with the cornea is not ideal for clinical translation and cannot be performed during the cross-linking procedure.
      Advances in both the sensitivity and imaging rate of phase-sensitive optical coherence tomography (PhS-OCT) have led to non-contact imaging systems that track propagating elastic waves launched by non-contact low-amplitude (less than a micrometer displacement) excitation.
      • Wang R.K.
      • Ma Z.
      • Kirkpatrick S.J.
      Tissue Doppler optical coherence elastography for real time strain rate and strain mapping of soft tissue.
      ,
      • Wang R.K.
      • Kirkpatrick S.
      • Hinds M.
      Phase-sensitive optical coherence elastography for mapping tissue microstrains in real time.
      Assuming an appropriate corneal mechanical model, dynamic OCE can measure in near real-time both the shape and biomechanical changes resulting from keratoconus and CXL surgery, with the potential for elastic moduli mapping at high spatial resolution.
      • Ambroziński Ł.
      • et al.
      Acoustic micro-tapping for non-contact 4D imaging of tissue elasticity.
      • Song S.
      • et al.
      Strategies to improve phase-stability of ultrafast swept source optical coherence tomography for single shot imaging of transient mechanical waves at 16 kHz frame rate.
      • Kirby M.A.
      • et al.
      Spatial resolution in dynamic optical coherence elastography.
      • Regnault G.
      • et al.
      Spatial resolution in optical coherence elastography of bounded media.
      Until recently, the existing body of literature did not present a clear picture on the appropriate elastic model for human cornea. Common models assumed an incompressible, isotropic, and linear elastic material to describe corneal deformation, where a single parameter, the Young’s modulus (E, or equivalently the shear modulus μ=E/3), defined elasticity. However, the in-plane collagen arrangement in the cornea leads to an anisotropic biomechanical response,
      • Pitre J.J.
      • et al.
      Nearly-incompressible transverse isotropy (NITI) of cornea elasticity: model and experiments with acoustic micro-tapping OCE.
      i.e. corneal in-plane elastic properties are different from those out-of-plane, which suggests that the cornea more closely resembles a transverse isotropic material. As a result, up to three orders of magnitude differences have been reported from mechanical tests between the in-plane Young’s modulus and the out-of-plane shear modulus depending on the method used.
      • Pitre J.J.
      • et al.
      Nearly-incompressible transverse isotropy (NITI) of cornea elasticity: model and experiments with acoustic micro-tapping OCE.

      Petsche, S. J., Chernyak, D., Martiz, J., Levenston, M. E. & Pinsky, P. M. Depth-Dependent Transverse Shear Properties of the Human Corneal Stroma. 53, (2017).

      • Hatami-Marbini H.
      Viscoelastic shear properties of the corneal stroma.
      • Hatami-Marbini H.
      • Etebu E.
      Hydration dependent biomechanical properties of the corneal stroma.
      Consequently, corneal elastic moduli obtained with isotropic models are highly inconsistent, greatly depend on the measurement technique and cannot be used to predict corneal deformation.
      • Hatami-Marbini H.
      Viscoelastic shear properties of the corneal stroma.
      ,
      • Bekesi N.
      • Dorronsoro C.
      • De La Hoz A.
      • Marcos S.
      Material properties from air puff corneal deformation by numerical simulations on model corneas.
      • Boschetti F.
      • Triacca V.
      • Spinelli L.
      • Pandolfi A.
      Mechanical characterization of porcine corneas.
      • Wollensak G.
      • Spoerl E.
      • Seiler T.
      Stress-strain measurements of human and porcine corneas after riboflavin-ultraviolet-A-induced cross-linking.
      • Zeng Y.
      • Yang J.
      • Huang K.
      • Lee Z.
      • Lee X.
      A comparison of biomechanical properties between human and porcine cornea.
      • Elsheikh A.
      • Alhasso D.
      Mechanical anisotropy of porcine cornea and correlation with stromal microstructure.
      • Hatami-Marbini H.
      • Jayaram S.M.
      Effect of UVA/riboflavin collagen crosslinking on biomechanics of artificially swollen corneas.
      • Lanchares E.
      • Del Buey M.A.
      • Cristóbal J.A.
      • Lavilla L.
      • Calvo B.
      Biomechanical property analysis after corneal collagen cross-linking in relation to ultraviolet A irradiation time.
      • Aslanides I.M.
      • et al.
      Assessment of UVA-Riboflavin Corneal Cross-Linking Using Small Amplitude Oscillatory Shear Measurements.
      • Søndergaard A.P.
      • Ivarsen A.
      • Hjortdal J.
      Corneal resistance to shear force after UVA-riboflavin cross-linking.
      Because the anisotropic cornea cannot be characterized with a single elastic modulus, more appropriate models are required.
      Recently, a model of a nearly incompressible transversely isotropic (NITI) medium was proposed based on corneal tissue microstructure.
      • Pitre J.J.
      • et al.
      Nearly-incompressible transverse isotropy (NITI) of cornea elasticity: model and experiments with acoustic micro-tapping OCE.
      It has been used to measure elastic anisotropy (i.e. in-plane, μ, and out-of-plane, G, shear moduli) in porcine cornea using non-contact acoustic micro-tapping OCE (AμT-OCE). Notably, AμT-OCE is the only reported method that can quantify both μ (and therefore E=3μ) and G simultaneously in a non-invasive, non-contact manner.
      • Kirby M.A.
      • et al.
      Delineating corneal elastic anisotropy in a porcine model using non-contact optical coherence elastography and ex vivo mechanical tests.
      This is critical for in vivo studies of human corneal physiology and pathophysiology.
      In this study, the NITI model was used to compute anisotropic mechanical moduli (E and G) from elastic wave fields generated and detected using AμT-OCE in both ex vivo untreated human donor corneas and in samples subject to riboflavin/UV collagen crosslinking (UV-CXL) at physiologically relevant controlled pressures. Mechanical moduli were then directly measured in one subgroup of the same samples using destructive mechanical methods that loaded tissue in tension (in the plane of stromal fibers in tensile extensometry) and in shear (between the lamellar sheets in shear rheometry) to independently quantify E and G. We show that the human cornea is highly anisotropic, and thus current isotropic models cannot predict corneal deformation under load.
      Corneal mechanical anisotropy was also probed using AμT-OCE on a set of samples before undergoing UV-CXL and immediately after the procedure, as well as on a set of samples scanned during CXL, to both demonstrate potential in-line monitoring of corneal stiffness intraoperatively and understand the range of cornea moduli changes induced by CXL. The non-contact nature of AμT-OCE makes it promising to study corneal disease progression through objective measurements of corneal moduli. Indeed, it can play a critical role in optimizing CXL treatment of progressive keratoconus.

      Methods & Materials

      Cornea Preparation

      A total of 22 human corneal-scleral rings stored in Optisol (Chiron Ophthalmics, Irvine, California, USA) were secured through both the University of Utah Lion’s Eye Bank (Murray, UT) and CorneaGen (Seattle, WA). Exclusion parameters included tissue over 2 months from enucleation, progressive corneal disease, and visual corneal damage. Such broad exclusion criteria were necessary to obtain research samples and resulted in a wide range of both ages and times between enucleation and analysis. Corneas were placed into three groups (Group A, B, and C) for different experiments based on their availability (Figure 1). The study adhered to the tenants in the Declaration of Helsinki and Institutional Review Board (IRB)/Ethics Committee approval was obtained.
      Figure thumbnail gr1
      Figure 1Schematic of the three different experiments performed and the corresponding number of samples in each group.
      In Group A, 10 samples were subject to both non-contact AμT-OCE and destructive mechanical testing (shear rheometry and tensile extensometry) to determine mechanical moduli in both normal and cross-linked corneas. The results obtained were analyzed using different methods to justify the use of the NITI model (described below) in human cornea. All samples were removed from solution upon acquisition and fitted to an artificial anterior chamber (AAC, Barron, CorzaMedical, Parsippany, New Jersey) connected via the inlet port to a bath filled with balanced saline solution (BSS). The outlet port was closed to induce hydrostatic pressure along the anterior corneal surface. The bath could be raised and lowered relative to the cornea surface (according to a height previously calibrated using a digital hydrostatic pressure sensor) to control the intraocular pressure (IOP) over a range of 5 mmHg – 20 mmHg, depending on the group under study. Each sample was inflated to a desired pressure for 5 minutes, over which a single drop of BSS was applied to prevent corneal dehydration. Corneal hydration was previously validated using this method, where a 50 μL dropper was used to apply BSS to the cornea every 5 minutes and the thickness remained unchanged.
      Corneas in group A were selected as untreated (group A1, n = 5) and cross-linked (group A2, n = 5) based on their availability. The average age of the untreated group (A1) was 53 ± 5 years and the samples were tested on average 31 ± 18 days from preservation. In the CXL group (A2), the average age was 29 ± 10 years and the samples were tested on average 15 ± 5 days from preservation.
      In group A1, each cornea was scanned with AμT-OCE at physiologically relevant pressures (5, 10, 15, and 20 mmHg). Immediately following AμT-OCE scanning, each sample was transported in a BSS-damped cloth for parallel-plate rheometry at room temperature before sectioned into strips approximately 6 mm wide and transported for tensile testing.
      In group A2, all samples were inflated to 15 mmHg and then subjected to a CXL procedure following the Dresden protocol prior to being scanned with AμT-OCE at pressures from 5 mmHg to 20 mmHg. Following CXL, corneas were transported for mechanical testing. Groups A1 and A2 independently demonstrated corneal anisotropy in both untreated cornea and those that underwent CXL.
      In group B, n = 7 cornea were inflated to 15 mmHg and mechanical moduli were measured on the same sample using non-contact AμT-OCE before and after CXL. The average age was 51 ± 17 years and samples were scanned on average 30 ± 5 days from preservation. Group B provided a paired analysis of biomechanical changes on the same samples resulting from the CXL procedure.
      In group C, n = 5 cornea were inflated to 15 mmHg and mechanical moduli were measured using AμT-OCE preoperatively, intraoperatively, and following CXL. The average age was 62 ± 8 years and samples were scanned on average 31 ± 15 days from preservation. Group C helped determine how quickly biomechanical changes occurred in each cornea during CXL.

      Collagen cross-linking (CXL) procedure

      The CXL procedure followed the widely adopted Dresden protocol.
      • Baiocchi S.
      • Mazzotta C.
      • Cerretani D.
      • Caporossi T.
      • Caporossi A.
      Corneal crosslinking: Riboflavin concentration in corneal stroma exposed with and without epithelium.
      ,
      • Wollensak G.
      • Spoerl E.
      • Seiler T.
      Riboflavin/ultraviolet-A-induced collagen crosslinking for the treatment of keratoconus.
      A 0.1% riboflavin (RF) solution in 20% dextran was first prepared using 500,000 weight dextran (Alfa Aesar, Ward Hill, Massachusetts, USA) in isotonic saline. Before applying this solution, the epithelium was debrided using a blunt knife. Next, a 50 μL drop of riboflavin solution was placed on the cornea every 2 minutes over a total duration of 30 minutes. After riboflavin loading, each cornea was exposed to 3 mW/cm2 of 370 nm (UV) light for 30 minutes. Riboflavin solution was reapplied every 5 minutes during the 30-minute irradiation period.
      For the samples in group A2 (CXL group), a baseline AμT-OCE scan was taken prior to and post epithelial debridement. The pre- and post- epithelial debridement measurements showed no statistical difference. As such, the post-epithelial debridement scan was used as the ‘pre-CXL’ dataset. For samples in group B, (pre- and post- CXL), the pre-CXL scan was taken immediately following epithelium removal and prior to applying riboflavin/dextran. All corneas were then scanned following 30 minutes of riboflavin/dextran solution exposure and 30-minute UV light exposure.
      Cornea samples in Group C were used to explore the rate of changes in cornea mechanical moduli during UV irradiation. In this experiment, a baseline scan was taken prior to and following epithelial debridement. As with Group A, the pre- and post- epithelial debridement measurements showed no statistical difference, and the post-epithelial debridement scan was considered the initial untreated dataset. Then, one 50μL drop of 0.1% RF in 20% dextran solution was placed on each cornea every 2 minutes for 30 minutes. Excess solution was removed, and a post-riboflavin scan was taken. Cornea were then exposed to 3 mW/cm2 of 370 nm UV light for 30 minutes, and drops were continued every 5 minutes during irradiation. Excess solution was removed from the tissue surface and the cornea was scanned every 90 seconds during the 30-minute UV exposure. Following UV exposure, 5 post-CXL scans were performed.

      Nearly-incompressible transverse isotropic (NITI) model of corneal elasticity

      To account for cornea collagen microstructure consisting of thin layers (lamellas) primarily oriented in-plane of the stroma, the macroscopic stress-strain relationship in the cornea for small deformations is described by Hook’s law in a NITI material, taking the form (in Voigt notation)
      • Kirby M.A.
      • et al.
      Delineating corneal elastic anisotropy in a porcine model using non-contact optical coherence elastography and ex vivo mechanical tests.
      :
      where σij denotes engineering stress, εij denotes engineering strain, τij denotes shear stresses, γij = 2εij denotes shear strains, and the subscripts x, y, and z refer to standard Cartesian axes. Thus, there are four independent elastic constants, λ, δ, G, and μ, defining the cornea’s elasticity matrix. Because the cornea is nearly-incompressible, the longitudinal modulus, λ, does not influence deformation. G and μ represent out-of-plane and in-plane shear moduli, respectively, whose magnitudes were shown to differ in corneal tissue. Indeed, several studies (including OCE,
      • Pitre J.J.
      • et al.
      Nearly-incompressible transverse isotropy (NITI) of cornea elasticity: model and experiments with acoustic micro-tapping OCE.
      ,
      • Kirby M.A.
      • et al.
      Delineating corneal elastic anisotropy in a porcine model using non-contact optical coherence elastography and ex vivo mechanical tests.
      and mechanical testing
      • Hatami-Marbini H.
      Viscoelastic shear properties of the corneal stroma.
      ,
      • Petsche S.J.
      • Chernyak D.
      • Martiz J.
      • Levenston M.E.
      • Pinsky P.M.
      Depth-dependent transverse shear properties of the human corneal stroma.
      ,
      • Ramirez-Garcia M.A.
      • Sloan S.R.
      • Nidenberg B.
      • Khalifa Y.M.
      • Buckley M.R.
      Depth-Dependent Out-of-Plane Young’s Modulus of the Human Cornea.
      ) have reported the out-of-plane shear modulus, G, to be on the order of tens of kPa, whereas the in-plane shear modulus, μ, to be on the order of MPa.
      • Lanchares E.
      • Del Buey M.A.
      • Cristóbal J.A.
      • Lavilla L.
      • Calvo B.
      Biomechanical property analysis after corneal collagen cross-linking in relation to ultraviolet A irradiation time.
      ,
      • Kirby M.A.
      • et al.
      Delineating corneal elastic anisotropy in a porcine model using non-contact optical coherence elastography and ex vivo mechanical tests.
      The in-plane Young’s modulus
      ET=3μ+μδ4μ+δ
      (2)


      is defined by μ and δ, but does not depend on G. In Refs.
      • Kirby M.A.
      • et al.
      Delineating corneal elastic anisotropy in a porcine model using non-contact optical coherence elastography and ex vivo mechanical tests.
      ,
      • Kirby M.A.
      • et al.
      Probing elastic anisotropy of human skin in vivo with light using non-contact acoustic micro-tapping OCE and polarization sensitive OCT.
      we showed, that
      2μ<δ<0
      (3)


      to obey cornea symmetry but highlight that δ cannot be extracted from guided wave propagation. Fortunately, constraints on δ limit the range of the in-plane Young’s modulus to
      2μ<ET<3μ.
      (4)


      Here we assume that
      E=ET3μ,
      (5)


      producing slight overestimates of actual values. Note that due to strong corneal anisotropy, δ likely varies little between corneas in a population. The exact relationship between μ and E can be clarified with tensile testing and is the focus of future studies. Even with this approximation (Eq. 5), the elastic moduli E and G can fully characterize corneal deformations for small deformations.

      Acoustic Micro-tapping Optical Coherence Elastography (AμT-OCE)

      Elastic moduli of human cornea were quantified using wave fields of guided mechanical waves generated and measured with a phase-sensitive spectral-domain OCT system combined with an air-coupled AμT ultrasound transducer. The OCT system used polarization-maintaining optical fibers and components, as described in detail previously.
      • Kirby M.A.
      • et al.
      Delineating corneal elastic anisotropy in a porcine model using non-contact optical coherence elastography and ex vivo mechanical tests.
      To generate and track propagating elastic waves, the OCT system operated in M-B-mode, in which a sequence of 512 A-scans were repeated at the same location, (referred to as an M-scan). A trigger signal was sent to a function generator at the beginning of each M-Scan to direct a 100 μs-long chirped (1 MHz-1.1 MHz) waveform to an air-coupled piezoelectric transducer (AμT source) with a matching layer, providing a temporally localized and spatially focused acoustic ‘push’.
      • Ambroziński Ł.
      • et al.
      Acoustic micro-tapping for non-contact 4D imaging of tissue elasticity.
      The acoustic force created a localized displacement (hundreds of nanometers in amplitude) at the cornea’s surface. The pulse/M-scan sequence was captured at 256 spatial locations across the imaging plane to create a 3-D volume (z, x, t). A complete M-B–scan consisted of 1024 depth × 256 lateral locations × 512 temporal frames (captured at a functional 46 kHz frame rate) with an effective imaging range of 1.5 mm × 10 mm x 11 ms (axial × lateral × temporal). Each scan took ∼3 seconds to acquire and save. Note that the OCT scanning range of 10 mm in the lateral direction was partially shaded (∼1.5 mm from the first scan location) by the AμT-OCE transducer.
      The resulting three-dimensional dataset was used to reconstruct central corneal structure pre- and post- CXL, as well as track propagating mechanical waves in the cornea following AμT excitation (Figure 2). The central corneal thickness pre- and post- CXL was calculated from the OCT structural image after correcting for the cornea’s refractive index (assumed to be 1.38) (Figure 3a,d). The OCT-measured local particle axial vibration velocity (vz(x,z,t)) was obtained from the optical phase difference Δφopt(x,z,t) between two consecutive A-line scans at each location.
      • Wang R.K.
      • Kirkpatrick S.
      • Hinds M.
      Phase-sensitive optical coherence elastography for mapping tissue microstrains in real time.
      Surface vibrations were detected using automatic segmentation of the anterior surface based on an edge detection algorithm. One half of a Gaussian window (HWHM = 90 μm, weight decreasing with depth) was applied as a weighted-average to phase data in the anterior 183 μm of the cornea, resulting in robust sampling of the vertical displacement for propagating guided elastic waves along the air-tissue boundary. Guided wave signals in the cornea recorded at different time moments were combined to form a wave field (Figure 3b,e).
      Figure thumbnail gr2
      Figure 2a) AμT-OCE imaging system with cornea inflated via artificial anterior chamber undergoing UV- CXL treatment. b) Schematic of mechanical wave propagation within the cornea. c) Snapshots of elastic transients detected at different time instants using phase-sensitive OCT.
      Figure thumbnail gr3
      Figure 3Central cornea cross-section and automatic segmentation to determine the surface location and thickness a) pre- and d) post-CXL. Wave fields of propagating guided waves in the same cornea b) pre- and e) post-CXL tracked over ∼8.5 mm of corneal tissue. Best-fit solution to the dispersion equation (based on a unique combination of elastic moduli, E and G, displayed in red) on top of the measured waveform in the 2-D Fourier spectrum for the corresponding cornea c) pre- and f) post-CXL.
      A temporal super Gaussian filter (SG) that followed the maximum vibration velocity of the wave-field tmwf(x) at each discrete position x, was then applied:
      SG(t)=exp[(12(ttmwf(x)σt)2)2](6)


      with σt = 0.5 ms. Eq. 6 acted as a moving temporal window around the peak of the space-time (x-t) plot to remove any unwanted reverberations around the main elastic wave-mode, as can be seen in Figure 3b,e. A two-dimensional Fourier transform was computed to describe wavefields in the frequency-wavenumber (f-k) domain (Figure 3c,f).
      A solution to the equation for guided wave propagation in the NITI material
      • Pitre J.J.
      • et al.
      Nearly-incompressible transverse isotropy (NITI) of cornea elasticity: model and experiments with acoustic micro-tapping OCE.
      was found that most closely matched the experimental f-k spectrum using an iterative routine that converged on a best-fit of the dispersion relation. By varying both the in-plane (μ, assuming tensile isotropy where μ=E/3) and out-of-plane, G, shear moduli and using the corneal thickness, h, as a constraint, a dispersion relation that closely matched experimental data was determined (see Supplemental Methods 1, 2 for details on the fitting routine). In all samples, the wave energy was mainly contained within the A0 mode
      • Kirby M.A.
      • et al.
      Optical coherence elastography in ophthalmology.
      ,
      • Pitre J.J.
      • et al.
      Nearly-incompressible transverse isotropy (NITI) of cornea elasticity: model and experiments with acoustic micro-tapping OCE.
      and fit to a theoretical solution of the A0 mode assuming a linear-elastic NITI material.
      Due to the nature of the weighted fitting routine, residuals are not generated, making traditional confidence interval methods difficult to apply. As such, asymmetrical error bars corresponding to uncertainties in E and G for each cornea (due to both model and data error) are described in Supplemental Methods 2. All cases that produced a poor fit quality (detailed in Supplemental Methods 3) were omitted.

      Destructive biomechanical testing

      Following AμT-OCE, sub-groups A1 and A2 were subject to mechanical testing under tensile and shear loading.
      The frequency-dependent shear behavior of a subset of corneal buttons was measured using a rheometer (Anton Paar MCR 301 Physica, Figure 4a) over a range of 0.16-16 Hz using a 5 N compressive preload and a peak shear strain of ∼.1%. Rheometry produced estimates of the frequency-dependent storage (G(ω)) and loss (G(ω)) components, corresponding to the real and imaginary parts of the modulus G.
      Figure thumbnail gr4
      Figure 4a) Cornea buttons prepared and c) loaded in parallel-plate rheometry to measure out-of-plane shear modulus, G. b) Cornea strips prepared and d) clamped during extension testing to determine in-plane Young’s modulus, E.
      Each sample was then cut into strips for tensile testing and pneumatically clamped (2752-005 BioPuls submersible pneumatic grips, 250 N max load) to the edge of the cornea (Figure 4b). A 50 mN pre-load was applied and each sample was stretched at 2 mm/min (Instron model 5543) up to either 10% strain, 10 N load, or corneal breakage (whichever occurred first). Two load-unload cycles were performed to precondition the tissue, followed by 3 rounds of force-elongation followed by relaxation. The force-elongation measurement was converted to stress-strain according to sample geometry. A second order exponential was fit using three-sets of raw data to determine the stress-strain curve. The in-plane Young’s modulus, E, was defined as the tangential slope of the stress-strain curve. Consequently, extension testing provided a value for the strain-dependent Young’s modulus, E, up to 10% strain.

      Results

      Elasticity changes with IOP in untreated and CXL cornea

      Corneal Thickness

      The mean thickness of all samples in Group A is presented in Table 1 for all tests. The CXL group (A2) was statistically thinner (p<.05) than the untreated group (A1) at all pressures tested with OCE and during rheometry. Corneal thickness in AμT-OCE was determined by OCT imaging; in rheometry - via parallel plate separation; and in extension testing - using a digital micrometer. Note that the OCE set-up placed the cornea under inflating forces and assumed a refractive index, rheometry required a flattening of the cornea and was insensitive to differences in the central and peripheral thickness, and the digital micrometer required hand-tuning.
      Table 1Corneal thickness for each testing condition.
      Mean thickness [μm](± standard deviation)
      OCERheometryExtension Testing
      5mmHg10mmHg15mmHg20mmHg
      Group A1 (Untreated)839 (±230)821 (±219)825 (±260)816 (±259)852 (±94)781 (±161)
      Group A2(CXL)630 (±114)633 (±120)631 (±123)633 (±128)654 (±107)645 (±129)
      Because OCE enabled measurements in cornea inflated to different IOP, a general (but not statistically significant) decrease in thickness was measured as pressure increased in the untreated group. Inflation induced thickness changes were not corrected for in rheometry and extension testing due to limitations in the experimental set up.

      Out-of-plane shear modulus, G

      In the ‘untreated’ group (A1), the OCE-determined out-of-plane shear modulus, G, for each sample is shown in Figure 5a. The statistical mean and standard deviation of the moduli between samples was 70 kPa (± 29 kPa) at 5 mmHg, 105 kPa (± 55 kPa) at 10mmHg, 169 kPa (± 129 kPa) at 15 mmHg, and 240 kPa (± 191 kPa) at 20 mmHg. In the ‘CXL’ group A2, the OCE-measured out-of-plane shear moduli, G, are also shown in Figure 5a, where the statistical mean and standard deviation of the moduli were 151 kPa (± 67 kPa) at 5 mmHg, 242 kPa (± 76 kPa) at 10 mmHg, 290 kPa (± 78 kPa) at 15 mmHg, and 334 kPa (± 99 kPa) at 20 mmHg.
      Figure thumbnail gr5
      Figure 5a) Out-of-plane shear modulus, G, in untreated (A1) and CXL (A2) cornea groups measured with AμT-OCE at intraocular pressures from 5-20 mmHg respectively. b) Shear storage modulus, G, measured with parallel plate rheometry for untreated (A1) and CXL (A2) cornea groups.
      A relative change of modulus, G, with increasing IOP for both untreated and CXL groups can be seen, where the mean shear modulus, G, increased at 20 mmHg (relative to the value at 5 mmHg) by a factor of 3.4 in the normal group compared with 2.2 in the CXL group.
      Mechanical measurements using parallel-plate rheometry are shown in Figure 5b for both untreated and cross-linked corneas. In untreated samples, the statistical mean and standard deviation of the moduli between samples were 201 kPa (± 55 kPa) at 0.16 Hz and 336 kPa ± (56 kPa) at 16 Hz. For the cross-linked group, the statistical mean and standard deviation were 270 kPa (±34 kPa) at 0.16 Hz and 388 kPa (±54 kPa) at 16 Hz. Rheometry measurements lie within the range of OCE-determined modulus, G.
      Individual results from each cornea have been included in Supplemental Methods 4. Note that OCE measured values in individual corneas generally tracked with rheometery measurements. Uncertainty intervals and detailed exclusion criterion can also be found in the Supplemental Methods.
      OCE-measured values of G were statistically different (p<0.05) between untreated and CXL groups at 5 mmHg and 10 mmHg, but not at 15 mmHg (p=.11) and 20 mmHg (p=.36). The means of rheometery results were not statistically different between groups at any frequency.

      In-plane Young’s modulus, E

      AμT-OCE experiments simultaneously evaluated in-plane, μ, and out-of-plane, G, shear moduli. The modulus μ can be then converted to the in-plane Young’s modulus (E=3μ).
      • Kirby M.A.
      • et al.
      Spatial resolution in dynamic optical coherence elastography.
      ,
      • Wollensak G.
      • Spoerl E.
      • Seiler T.
      Riboflavin/ultraviolet-A-induced collagen crosslinking for the treatment of keratoconus.
      Note that E cannot be determined from rheometry measurements. We used tensile extensometry to compare E with that determined from AμT-OCE.
      In the untreated group (A1), the OCE-measured in-plane Young’s modulus, E, for each sample is shown in Figure 6a. The statistical mean and standard deviation of the moduli between samples were 15 MPa (± 5 MPa) at 5mmHg, 21 MPa (± 8 MPa) at 10 mmHg, 24 MPa (± 9 MPa) at 15 mmHg, and 24 MPa (± 13 MPa) at 20 mmHg. In the CXL group (A2), each of the in-plane Young’s moduli, E, is also shown in Figure 6a, where the statistical mean and standard deviation of the moduli between samples were 37 MPa (± 7 MPa) at 5 mmHg, 39 MPa (± 7 MPa) at 10 mmHg, 44 MPa (± 6 MPa) at 15 mmHg, and 47 MPa (± 7 MPa) at 20 mmHg.
      Figure thumbnail gr6
      Figure 6a) In-plane Young’s modulus, E, in untreated group (A1) and CXL group (A2) measured with AμT-OCE at intraocular pressures from 5-20 mmHg. b) Strain-dependent Young’s moduli measured via extension testing up to 10% strain in untreated group (A1) and CXL group (A2). E was determined via extension testing to 10% strain, or where visible tissue damage occurred. The dashed lines correspond with plus or minus one standard deviation. Individual tests can be found in the Supplemental Methods.
      Relative changes in Young’s modulus with IOP were smaller for CXL corneas compared to that for untreated ones where Young’s modulus for untreated corneas increased by a factor of 1.6 with an increase in IOP from 5 mmHg to 20 mmHg and Eincreased by ∼1.3 for CXL cornea.
      The in-plane Young’s modulus determined by extension testing at 1-10% strains following dissection of the samples is shown in Figure 6b, where the statistical mean and standard deviation at 1% strain were 3.2 MPa (±2.2 MPa) and at 10% strain were 51 MPa (±31 MPa). In the CXL group (A2), the in-plane Young’s modulus determined by extension testing at 1-10% strains is also shown in Figure 6b where the statistical mean at 1% strain was 2.9 MPa (±0.9 MPa). The high strain modulus was either at 10% strain, or at the inflection point before structural damage occurred. The high strain modulus in the CXL group was 74 MPa (±50 MPa). Samples in the CXL group demonstrated significant tissue damage at higher strains and required more force to extend.
      OCE-measured values of E were statistically different between groups at all pressures (p<0.1 for IOP of 5 mmHg to 20 mmHg, respectively.) The means of the best fit curve of the strain dependent tangential modulus were not statistically different across all strains. Uncertainty intervals and individual results from each cornea have also been included in Supplemental Methods 4.
      The independent sub-group study suggests that in general, (i) both moduli (G and E) could be measured in ex vivo human cornea using the NITI model in untreated and CXL treated samples (ii) OCE measurements generally agree with mechanical tests. Thus, the effect of CXL on corneal elasticity can be explored in more detail without direct comparison to mechanical tests. This result also suggests that OCE can monitor elasticity changes throughout the entire treatment cycle, including untreated corneas pre-, intra- and post-CXL.
      Additionally, the results suggest that one-to-one comparisons before and after CXL are required to better understand the effect of UV-CXL. Such scans can measure CXL-induced changes for each individual cornea rather than relying on group comparisons between untreated and treated samples. Analyzing results for every cornea independently should greatly reduce the effect of population variability of corneal properties, which is key to developing personalized models of the eye for optimal surgery planning. Results on ex vivo preliminary studies are considered in the sub-sections below.

      Independent monitoring of corneal moduli pre- and post- CXL

      Results for each sample in Group B, where moduli were quantified immediately before and after cross-linking, can be seen in Figure 7a and Figure 7b. Clearly, changes induced by CXL vary from cornea to cornea, suggesting that individual treatment plans may be required to reach desired moduli outcomes.
      Figure thumbnail gr7
      Figure 7CXL-induced differences in a) in-plane Young’s modulus, E, and b) out-of-plane shear modulus, G, for cornea in group B. Error bars correspond with uncertainty intervals described in the Supplemental Methods. Note that in one sample (cornea #3) all 5 scans demonstrated high uncertainty in G following CXL, thus the value is displayed as an ‘open’ triangle and omitted from analysis of the means. c) Thickness for each individual cornea pre- CXL and post- CXL.
      The statistical mean (standard deviation) of E changed from 19 MPa (± 4 MPa) to 43 MPa (±11 MPa) and the shear modulus, G, from 188 kPa (±148 kPa) to 673 kPa (± 440 kPa) (see Figure 8). Note that in cornea #3, (post-CXL), (Figure 7b), all 5 repeat scans demonstrated high uncertainty in G. As such, its value was omitted from the calculation of mean values.
      Figure thumbnail gr8
      Figure 8The mean and standard deviation in a) G and b) Eprior to and following UV-CXL in n=5 samples.
      While both moduli increased following UV exposure in RF/dextran-soaked cornea, the average change in in-plane Young’s modulus, E, was approximately double (2.3), whereas the out-of-plane shear modulus G changed by more than three times (3.6). Notably, both E and G increased for all samples following CXL.
      Both E and G had significantly different stiffening responses to CXL, where statistical significance (p < 0.05) was determined using a two-tailed t-test where the null hypothesis was no difference between the means in the pre- and post -CXL values. The p-value for G was 0.03 and for E was 0.02, indicating a significant stiffening response in both moduli.
      The central corneal thickness following epithelium removal was calculated from the OCT structural image pre- and post- CXL. The measured thickness pre-CXL was 577 μm (± 58 μm) and reduced to 450 μm (± 80 μm) post-CXL. The measured thickness was also statistically different in the samples following CXL (p = 0.001).

      Evaluation of both corneal moduli during CXL procedure

      To demonstrate the ability of AμT-OCE to directly quantify corneal biomechanics intraoperatively, E, G, and thickness were measured in group C every 90 seconds during the procedure. A moving average on both the output value and uncertainty intervals was performed over 4.5 minutes (the full data-set has been included in the Supplemental Materials). The mean and standard deviation of the relative modulus (i.e., modulus at any time normalized to the modulus pre-CXL) can be seen in Figure 9, along with a linear fit (over time) to highlight individual changes in G, E, and central cornea thickness with CXL. While both moduli increased with UV exposure, E changed on average by about 2 times (from 15 MPa (±4 MPa) to 31 MPa (± 6 MPa)) whereas G changed by approximately 4 times (from 107 kPa (±26 kPa) to 421 kPa (±110 kPa)) following CXL. Both moduli changed significantly (p<.01) between pre- and post-CXL values. Each cornea changed stiffness and thickness at a slightly different rate. Non-negligible stiffening also occurred from the RF/dextran alone within the first 30 minutes. Note that two (2) of the corneas resulted in a poor fit for G over the last 6 minutes. Four (4) of the 5 samples provided sufficient fit for G in the scans taken immediately following CXL.
      Figure thumbnail gr9
      Figure 9a) Young’s Modulus E and b) Shear Modulus G, relative to the pre-CXL value; c) central cornea thickness relative to the pre-CXL value. Black markers are the mean and standard deviation of the relative changes between the n=5 samples. The solid colored lines represent a linear best-fit of the relative change from each individual cornea. For 2 of the corneas, the fit for G was very poor over the last 6 minutes, so the values have been omitted from the displayed average. The post CXL average includes 4 corneas.
      The cornea progressively thinned during crosslinking. The measured thickness pre- and post- CXL was 560 μm ± 70 μm and 340 μm± 20μm, respectively. Combined elasticity and thickness changes during CXL will have significant impact on both the overall corneal stiffness and potential refractive changes. Although this experiment required a slight modification of the FDA-approved Dresden protocol (i.e. removal of RF/dextran from the corneal surface prior to each scan), the results indicate that AμT-OCE potentially can monitor treatment progression.

      Discussion

      Mechanical anisotropy was confirmed in ex vivo human cornea, where the in-plane tensile (Young’s) modulus, E, was on average multiple orders of magnitude larger than the out-of-plane shear modulus, G. Similar to porcine cornea,
      • Kirby M.A.
      • et al.
      Delineating corneal elastic anisotropy in a porcine model using non-contact optical coherence elastography and ex vivo mechanical tests.
      the results of Group A suggest that human cornea can be approximated as a nearly incompressible transversely isotropic (NITI) material. Using the NITI model to describe corneal deformation plays a key role in explaining the order(s) of magnitude difference in corneal stiffness estimates extracted from shear- and tensile-based mechanical measurements.
      Guided elastic wave propagation mixes fundamental modes allowing E and G to be measured simultaneously in the cornea.
      • Pitre J.J.
      • et al.
      Nearly-incompressible transverse isotropy (NITI) of cornea elasticity: model and experiments with acoustic micro-tapping OCE.
      As shown previously for strongly anisotropic materials, a change in E/μ results in a small, but detectable change in A0-mode dispersion, particularly in the low frequency region (where the wavelength is long compared to the thickness). As such, the wave velocity is roughly dominated by the shear modulus, G, and applying an isotropic model to fit the same recorded waveforms of guided mechanical waves would lead to a rough approximation of G, not E or μ. In this case, Young’s modulus, E, calculated using the expression E=3G (as done previously
      • Tanter M.
      • Touboul D.
      • Gennisson J.L.
      • Bercoff J.
      • Fink M.
      High-resolution quantitative imaging of cornea elasticity using supersonic shear imaging.
      • Han Z.
      • et al.
      Optical coherence elastography assessment of corneal viscoelasticity with a modified Rayleigh-Lamb wave model.
      • Zvietcovich F.
      • et al.
      Longitudinal shear waves for elastic characterization of tissues in optical coherence elastography.
      ) would be highly underestimated compared to the correct relationship of E with μ. Such an underestimated E modulus used in prediction models would produce an inaccurate estimate of cornea stiffness, its deformation, and would fail to evaluate clinical procedure outcomes, including that for CXL.
      To better understand the limitations in quantifying moduli, a method to determine uncertainty intervals in OCE-measured modulus was presented in Supplemental Methods 2 using a goodness-of-fit routine that compared the A0 structure for different values of E, and G, independently. As demonstrated previously, the A0 mode is increasingly insensitive to changes in E as the degree of mechanical anisotropy increases (E >>G).
      • Pitre J.J.
      • et al.
      Nearly-incompressible transverse isotropy (NITI) of cornea elasticity: model and experiments with acoustic micro-tapping OCE.
      This suggests that when E is much larger than G, it can be very difficult to accurately determine E for typical SNR’s achieved with OCE. Signal averaging or multi-channel scanning may be used to improve SNR and enable more accurate determination of E, but can be difficult to perform in vivo. To predict refractive changes and model the cornea as a curved lens system, both G and Eare required, yet, one question that remains is the precision needed to measure a change in E corresponding to a physical change in corneal shape. Static models are required to answer this question and remain an area of future interest.
      The results of Group A suggest that AμT-OCE can quantify both G and E, with less certainty around E (Supplemental Methods 2). Although E in the NITI model can have a high degree of uncertainty, analyzing guided wave propagation remains the only non-contact, non-destructive method reported to date to estimate in-plane tensile stiffness.
      While elastic moduli measured via OCE and mechanical methods in this study generally agree, there are subtle experimental differences that make it difficult to quantitatively compare values between methods. For example, an artificial anterior chamber (AAC) was used to inflate corneal samples, providing additional boundary conditions in the lateral direction limiting the propagation distance of the guided wave over the cornea, that differ from those of both in vivo and destructive ex-vivo settings. While probing tissue inflated via an AAC is more like cornea loading in vivo, mechanical boundary conditions are markedly different from those for whole-globe samples. Together with AμT excitation, it can lead to modifications in the low-frequency region of the f-k spectrum, worsening fitting accuracy. Additionally, to facilitate extension testing, a portion of the sclera was left on each sample during rheometry, which was shown to induce a measurable increase in the apparent modulus G.
      • Kirby M.A.
      • et al.
      Delineating corneal elastic anisotropy in a porcine model using non-contact optical coherence elastography and ex vivo mechanical tests.
      Sample transportation, time limitations, and measurement precision also contributed to differences in measured thickness between OCE and destructive methods, potentially influencing final quantitative moduli values.
      In Group B, both the out-of-plane shear modulus, G, and in-plane Young’s modulus, E, increased significantly following cross-link formation. The measured increase in both G and E suggests that CXL strengthened the tensile and flexural response, as well as the ability of the cornea to resist shearing forces. The measured increases in G and E of approximately 4-fold and 2-fold, respectively, generally agreed with values previously reported. For example, Marbini et. al.
      • Hatami-Marbini H.
      • Rahimi A.
      Stiffening effects of riboflavin/UVA corneal collagen cross-linkingis hydration dependent.
      and Del Buey et. al.
      • Del Buey M.A.
      • et al.
      Immediate effect of ultraviolet - A collagen cross-linking therapy on the biomechanics and histology of the human cornea.
      reported a post-CXL increase in corneal Young’s modulus, E, by a factor of ∼1.5-2 in the 7-10% strain range, while Wollensak et al. report an increase of a factor of 4.5 (1.3 MPa- 5.9 MPa at 6% strain) in human cornea tested via strip extension within 1 hour of enucleation.
      • Wollensak G.
      • Spoerl E.
      • Seiler T.
      Stress-strain measurements of human and porcine corneas after riboflavin-ultraviolet-A-induced cross-linking.
      The results of this study are also consistent with Aslanides et. al.,
      • Aslanides I.M.
      • et al.
      Assessment of UVA-Riboflavin Corneal Cross-Linking Using Small Amplitude Oscillatory Shear Measurements.
      who reported an increase as high as 20-fold in the shear modulus G in porcine cornea following CXL (as measured using a parallel-plate scheme) and Sondergaard et. al.
      • Søndergaard A.P.
      • Ivarsen A.
      • Hjortdal J.
      Corneal resistance to shear force after UVA-riboflavin cross-linking.
      who reported a general 2-5 fold increase in G in human cornea under 5% compressive strain, where samples were probed using a parallel plate arrangement to provide a translational shear (as opposed to rotational).
      Following UV-CXL, the out-of-plane shear modulus, G, and the in-plane Young’s modulus, E, increased from untreated values by different multiples. Specifically, the shear modulus, G, (most likely related to the proteoglycan mesh and interlamellar connectivity) changed almost four times in CXL corneas compared to untreated corneas, whereas the tensile modulus, E, changed only about two times. This result suggests that CXL changes elastic properties of the connective tissue mesh between the lamellar sheets in addition to that of fibers, making the connective tissue or proteoglycan mesh apparently ‘stickier’. The effect of the connective matrix on mechanical stability is well known in different material science and engineering areas, where fibers embedded in a composite material can produce almost equivalent tensile properties as the fibers, with shear properties defined independently by the connective matrix. Consistent with this effect, the in-plane tensile properties of CXL cornea increased relatively less than the apparent connectivity between fibers.
      To date, the specific relationship between corneal micro- and macro-structure with shape and focusing power remains speculative. Based on the NITI model, it can be hypothesized that a loss in lamellar inter-connectivity or degradation of the proteoglycan host-mesh would produce a higher degree of lamellar slippage while degradation in fiber strength along its axis would produce reduced tensile strength and a corresponding change in shape. In keratoconus progression, it has been shown that reduced interlamellar branching can be found in the anterior portion that corresponds to a measurable reduction in tensile modulus.
      • Winkler M.
      • et al.
      Nonlinear optical macroscopic assessment of 3-D corneal collagen organization and axial biomechanics.
      Reduced biomechanical stability due to lamellar slippage and collagen degradation has been suggested as a potential mechanism behind disease progression,
      • Dupps W.J.
      • Wilson S.E.
      Biomechanics and wound healing in the cornea.
      ,

      Smolek, M. K. & Klyce, S. D. Is Keratoconus a True Ectasia? 118, (2000).

      • Hayes S.
      • Boote C.
      • Tuft S.J.
      • Quantock A.J.
      • Meek K.M.
      A study of corneal thickness, shape and collagen organisation in keratoconus using videokeratography and X-ray scattering techniques.
      • Fan R.
      • Chan T.C.Y.
      • Prakash G.
      • Jhanji V.
      Applications of corneal topography and tomography: a review.
      but remains difficult to monitor in a clinical setting. Still, studies of micro- and macro- biomechanical and biochemical stability of the cornea following CXL are required to better understand both short- and long-term effects on refractive changes.
      Due to difficulties in acquiring human research tissue, all samples were scanned once the viable donor period had passed (greater than 1-2 weeks following enucleation). As such, biomechanical changes due to tissue necrosis and swelling may have occurred in samples prior to testing. Note that the apparent corneal stiffness (both E and G) for different hydration levels has been explored in detail using numerous animal models.
      • Hatami-Marbini H.
      • Etebu E.
      Hydration dependent biomechanical properties of the corneal stroma.
      ,
      • Søndergaard A.P.
      • Ivarsen A.
      • Hjortdal J.
      Corneal resistance to shear force after UVA-riboflavin cross-linking.
      ,
      • Petsche S.J.
      • Chernyak D.
      • Martiz J.
      • Levenston M.E.
      • Pinsky P.M.
      Depth-dependent transverse shear properties of the human corneal stroma.
      ,
      • Hatami-Marbini H.
      • Rahimi A.
      The relation between hydration and mechanical behavior of bovine cornea in tension.
      • Seiler T.G.
      • Shao P.
      • Frueh B.E.
      • Yun S.H.
      • Seiler T.
      The influence of hydration on different mechanical moduli of the cornea.
      • Kazaili A.
      • Geraghty B.
      • Akhtar R.
      Microscale assessment of corneal viscoelastic properties under physiological pressures.
      In the present study, cross-linking via the mock Dresden protocol induced a change in thickness from approximately 577 μm to 450 μm, and a measured doubling of the Young’s modulus. The doubling in tensile modulus was greater than what would be expected due to swelling alone. As reported previously, the shear modulus G appeared less sensitive to hydration compared to the tensile modulus. In the samples tested in Group B, we measured a 3-fold increase in the shear modulus G; again, a change unlikely to result from thinning alone.
      While carefully accounting for swelling in all samples remains an area of future study, the degree to which swelling alone, versus cross-linking alone, changes mechanical properties remains unclear. Of course, it remains likely that the reported differences in stiffness were due to a combination of simultaneous cross-linking and thinning. Because the present study sought to demonstrate that non-contact OCE can quantify changes induced in corneas undergoing a procedure similar to a clinical protocol, it does not measure the effect of cross-linking alone. Instead, a tool to measure expected changes during and following CXL is presented. Note that the measured differences in thickness generally agree with what has been reported elsewhere in both in vivo and ex vivo studies using a similar CXL procedure.
      • Hatami-Marbini H.
      • Etebu E.
      Hydration dependent biomechanical properties of the corneal stroma.
      ,
      • Kontadakis G.A.
      • et al.
      In vitro effect of corneal collagen cross-linking on corneal hydration properties and stiffness.
      • Wollensak G.
      • Aurich H.
      • Pham D.T.
      • Wirbelauer C.
      Hydration behavior of porcine cornea crosslinked with riboflavin and ultraviolet A.
      • Hatami-Marbini H.
      • Rahimi
      A. Interrelation of Hydration, Collagen Cross-Linking Treatment, and Biomechanical Properties of the Cornea.
      Ideally, real-time monitoring of corneal stiffness during CXL would enable immediate feedback on procedure progression. In this study (Group C), we noted different rates of stiffening for individual corneas. We also note that there was a measurable difference in both moduli and thickness following the application of the riboflavin/dextran solution. Simply soaking the cornea in dextran can result in a measurable change in corneal thickness and stiffness,
      • Hatami-Marbini H.
      • Etebu E.
      Hydration dependent biomechanical properties of the corneal stroma.
      consistent with the pre-CXL and pre-UV measurements in Group C. As corneas were exposed to UV illumination, OCE could measure the overall effect of the mock CXL procedure (from both Rf/Dextran soaking and UV-illumination) in the Dresden protocol. If the effects of both total stiffening and the rate of stiffening induced by Rf/Dextran and UV-illumination need to be explored independently, further studies are required.
      Because research tissue was difficult to acquire, the inclusion of older aged samples (Supplemental Methods 6), may have resulted in generally higher moduli.
      • Elsheikh A.
      • et al.
      Assessment of corneal biomechanical properties and their variation with age.
      In fact, corneas from eye-banks tend to be from older donors, which makes it difficult to study samples from the age groups that most commonly present progressing keratoconus (i.e., age groups most likely to undergo CXL).
      • Alenezi B.
      • Kazaili A.
      • Akhtar R.
      • Radhakrishnan H.
      Corneal biomechanical properties following corneal cross-linking: Does age have an effect?.
      One issue with that remains a limitation in this study, for example, is that the mean age in group A1 was approximately 24 years greater than A2, suggesting that the pre-treatment moduli may be skewed toward higher values in group A1. Still, the younger tissues (group A2) demonstrated a relatively higher modulus post-CXL compared to the older (A2) samples. Note that Groups B and C were closer in age to group A1, and older than the age where keratoconus most commonly presents. Due to a host of microstructural differences and unique cross-linking behaviors between human and animal models
      • Wollensak G.
      • Spoerl E.
      • Seiler T.
      Stress-strain measurements of human and porcine corneas after riboflavin-ultraviolet-A-induced cross-linking.
      ,
      • Zeng Y.
      • Yang J.
      • Huang K.
      • Lee Z.
      • Lee X.
      A comparison of biomechanical properties between human and porcine cornea.
      ,
      • Elsheikh A.
      • Alhasso D.
      • Rama P.
      Biomechanical properties of human and porcine corneas.
      • Kohlhaas M.
      • et al.
      Biomechanical evidence of the distribution of cross-links in corneastreated with riboflavin and ultraviolet a light.
      • Scott McCall A.
      • et al.
      Mechanisms of corneal tissue cross-linking in response to treatment with topical riboflavin and Long-Wavelength Ultraviolet Radiation (UVA).
      • Beshtawi I.M.
      • O’Donnell C.
      • Radhakrishnan H.
      Biomechanical properties of corneal tissue after ultraviolet-A-riboflavin crosslinking.
      , even older human samples can provide valuable information and help demonstrate the potential for in vivo human use. Difficulties acquiring human tissues for research are not unique and have been recognized previously as a limiting factor in acquiring sufficient sample sizes.
      • Alenezi B.
      • Kazaili A.
      • Akhtar R.
      • Radhakrishnan H.
      Corneal biomechanical properties following corneal cross-linking: Does age have an effect?.
      ,
      • Kohlhaas M.
      • et al.
      Biomechanical evidence of the distribution of cross-links in corneastreated with riboflavin and ultraviolet a light.
      ,
      • Beshtawi I.M.
      • O’Donnell C.
      • Radhakrishnan H.
      Biomechanical properties of corneal tissue after ultraviolet-A-riboflavin crosslinking.
      While the moduli quantified with AμT-OCE are generally consistent with both literature and mechanical testing, the proposed method has several limitations. One is that depth-dependent moduli cannot be measured using an approach based on guided mechanical waves and, consequently, the measurements most likely correspond to depth-averaged values of corneal moduli. It has been shown that CXL is most dramatic in the anterior portion of the cornea,
      • Kohlhaas M.
      • et al.
      Biomechanical evidence of the distribution of cross-links in corneastreated with riboflavin and ultraviolet a light.
      which contains a complex fiber arrangement that likely determines most of its flexural (tensile) strength. It has also been shown that the relative increase in G is much greater in the anterior portion of the cornea.
      • Petsche S.J.
      • Chernyak D.
      • Martiz J.
      • Levenston M.E.
      • Pinsky P.M.
      Depth-dependent transverse shear properties of the human corneal stroma.
      Thus, if there was a significant increase in anterior cross-linking, much of the stroma may still be susceptible to inter-lamellar slippage. Exploring the importance of depth-resolved changes, as well as the interplay between moduli and intraocular pressure, remain directions of future interest.
      Treatment success in CXL is currently defined by secondary measures such as corneal shape and refractive power changes over time.
      • Hersh P.S.
      • et al.
      United States Multicenter Clinical Trial of Corneal Collagen Crosslinking for Keratoconus Treatment.
      To truly optimize outcomes, elastic moduli maps in cornea before and after surgery could evaluate individual stiffness correction requirements and monitor whether the required correction had been reached. As focal biomechanical weakening in the cone can occur in progressive keratoconus, localized biomechanical mapping is more likely to provide the information needed to develop a personalized model for each individual cornea suitable for screening, surgical planning, and treatment monitoring. Because elastic moduli maps may be combined with topography measurements provided by ophthalmic systems, topographic and mechanical changes would not only provide potentially important diagnostic information but may also be used to monitor localized cross-linking treatments. Although OCT can provide micron-scale resolution in tracking mechanical waves, the resultant OCE resolution in mapping mechanical properties across the cornea is lower.
      • Kirby M.A.
      • et al.
      Spatial resolution in dynamic optical coherence elastography.
      ,
      • Regnault G.
      • et al.
      Spatial resolution in optical coherence elastography of bounded media.
      Cornea is a bounded material and, therefore, guided mechanical waves further degrade the resolution in mapping moduli. While high resolution images of local group velocity have been demonstrated in 3-dimensional corneal sections,
      • Ambroziński Ł.
      • et al.
      Acoustic micro-tapping for non-contact 4D imaging of tissue elasticity.
      local mapping of elastic moduli has not yet been demonstrated. Methods improving spatial resolution in dynamic OCE measurements of bounded material (such as the cornea) are currently under investigation, and the demonstration of spatially resolved maps of cornea mechanical moduli is a subject of future studies.
      The standard Dresden protocol involves a 30-min corneal exposure to 370 nm UV radiation at an irradiance of 3 mW/cm2 following riboflavin/dextran saturation.
      • Wollensak G.
      • Spoerl E.
      • Seiler T.
      Riboflavin/ultraviolet-A-induced collagen crosslinking for the treatment of keratoconus.
      Several other protocols have been proposed which reduce exposure time. For example, accelerated CXL uses higher irradiance UV-A to shorten the time needed to deliver the equivalent total energy dose.
      • Hashemi H.
      • et al.
      Long-term Results of an Accelerated Corneal Cross-linking Protocol (18 mW/cm2) for the Treatment of Progressive Keratoconus.
      While different protocols have demonstrated varying degrees of stiffening based on the time of UV irradiance,
      • Lanchares E.
      • Del Buey M.A.
      • Cristóbal J.A.
      • Lavilla L.
      • Calvo B.
      Biomechanical property analysis after corneal collagen cross-linking in relation to ultraviolet A irradiation time.
      their translation remains limited and cannot be personalized without quantitative measurements of corneal elasticity, which is critical in guiding CXL treatment and predicting outcomes.
      Finally, we note that thickness remains an important parameter in determining whether or not to perform CXL based on the hypothesis that riboflavin serves as both a cross-linking agent and a means to protect the posterior segment of the eye from UV radiation.

      U.S. Department of Health Education And Welfare. Ocular Ultraviolet Effects from 295nm to 400nm in the Rabbit Eye. NIOSH Research Report (1977).

      • Caruso C.
      • et al.
      Customized Corneal Cross-Linking- A Mathematical Model.
      • Friedman M.D.
      • et al.
      Advanced corneal cross-linking system with fluorescence dosimetry.
      • Seiler T.G.
      • et al.
      Two-photon fluorescence microscopy for determination of the riboflavin concentration in the anterior corneal stroma when using the dresden protocol.

      Schumacher, S., Mrochen, M., Wernli, J., Bueeler, M. & Seiler, T. Optimization Model for UV-Riboflavin Corneal Cross-linking. 762–769 (2012). doi:10.1167/iovs.11-8059

      • Spoerl E.
      • Mrochen M.
      • Sliney D.
      • Trokel S.
      • Seiler T.
      Safety of UVA-Riboflavin Cross-Linking of the Cornea.
      • Wollensak G.
      • Spoerl E.
      • Reber F.
      • Seiler T.
      Keratocyte cytotoxicity of riboflavin/UVA-treatment in vitro.
      • Wollensak G.
      • Spoerl E.
      • Wilsch M.
      • Seiler T.
      Endothelial cell damage after riboflavin-ultraviolet-A treatment in the rabbit.
      While analytical models describing riboflavin loading, UV light exposure, and polymerization have been used to determine safety limits and predict surgical outcomes, current treatment parameters are designed assuming that a large majority of the UV light is absorbed or scattered in the anterior 250 to 350 μm of the corneal stroma in all patients. Thus, a minimum stromal thickness of 400 μm is required to protect corneal endothelium and deeper ocular structures.
      • Spoerl E.
      • Mrochen M.
      • Sliney D.
      • Trokel S.
      • Seiler T.
      Safety of UVA-Riboflavin Cross-Linking of the Cornea.
      This approach often excludes advanced keratoconus patients. We hypothesize that OCE can be used to design patient-specific procedures balancing the benefit between UV-radiation time and desired biomechanical stabilization, opening the possibility for optimized CXL treatment of even advanced keratoconus patients.
      In this study, we demonstrate the potential of AμT-OCE to monitor stiffening and thinning of the cornea intraoperatively, with varying rates of change in E and G. Although we demonstrated the potential to quantify both elastic moduli non-invasively using a non-contact near real-time procedure, it is important to note that results obtained in ex vivo studies (such as the one reported herein) may not fully correspond to changes expected in in vivo corneas using the same protocol. Given the favorable ectasia-stabilizing results with cornea cross-linking therapy, efforts aimed at optimizing the procedure’s treatment time, intra- and postoperative comfort, and efficacy remain an area of focus.

      Conclusions

      Corneal cross-linking is increasingly being used clinically to biomechanically stabilize the cornea in vision degrading diseases such as keratoconus or post-LASIK ectasia. While clinical outcomes are largely positive, there remains a gap in both understanding the mechanisms behind corneal stabilization and providing clinicians feedback on how those mechanisms may relate to disease progression and treatment plans.
      In this study, AμT-OCE quantified changes in both the in-plane Young’s modulus, E, and out-of-plane shear modulus, G, of human donor cornea due to UV induced crosslinking in a non-contact, non-destructive manner. The results demonstrated a significantly different stiffening response in both E and G, where G experienced a larger relative increase from the cross-linking procedure. This suggests that while CXL strengthened the tensile and flexural strength of the tissue, the ability of the cornea to resist shearing forces was more dramatically altered by the procedure.
      AμT-OCE provides a unique non-invasive and non-contact tool that can monitor and evaluate corneal elasticity, including changes induced by UV-CXL. CXL-induced changes in mechanical anisotropy likely carry diagnostic and prognostic information that warrant future studies.

      Acknowledgements

      The authors wish to thank Kit Hendrickson for assistance with figures.

      References

        • Meek K.M.
        • Boote C.
        The organization of collagen in the corneal stroma.
        Exp. Eye Res. 2004; 78: 503-512
        • Koudouna E.
        • et al.
        Evolution of the vertebrate corneal stroma.
        Prog. Retin. Eye Res. 2018; 64: 65-76
        • Borcherding M.S.
        • et al.
        Proteoglycans and collagen fibre organization in human corneoscleral tissue.
        Exp. Eye Res. 1975; 21: 59-70
        • Scott J.E.
        • Thomlinson A.M.
        The structure of interfibrillar proteoglycan bridges ('shape modules’) in extracellular matrix of fibrous connective tissues and their stability in various chemical environments.
        J. Anat. 1998; 192: 391-405
        • Meek K.M.
        • Knupp C.
        Corneal structure and transparency.
        Prog. Retin. Eye Res. 2015; 49: 1-16
        • Gandhi S.
        • Jain S.
        The anatomy and physiology of cornea.
        Keratoprostheses Artif. Corneas Fundam. Surg. Appl. 2015; 37: 19-25
        • Kotecha A.
        What Biomechanical Properties of the Cornea Are Relevant for the Clinician.
        Surv. Ophthalmol. 2007; 52: 109-114
        • Dupps W.J.
        • Wilson S.E.
        Biomechanics and wound healing in the cornea.
        Exp. Eye Res. 2006; 83: 709-720
        • Andreassen T.T.
        • Hjorth Simonsen A.
        • Oxlund H.
        Biomechanical Properties of Keratoconus and Normal Corneas.
        Exp. Eye Res. 1980; 31: 435-441
        • Chan E.
        • Snibson G.R.
        Current status of corneal collagen cross-linking for keratoconus: A review.
        Clin. Exp. Optom. 2013; 96: 155-164
        • Santhiago M.R.
        • Randleman J.B.
        The biology of corneal cross-linking derived from ultraviolet light and riboflavin.
        Exp. Eye Res. 2021; 202
        • De Moraes C.
        Risk factors for visual field progression in treated glaucoma.
        Arch. Ophthalmol. 2011; 129: 878
        • Roberts C.
        Concepts and misconceptions in corneal biomechanics.
        J. Cataract Refract. Surg. 2014; 40: 862-869
        • Eliasy A.
        • et al.
        Determination of Corneal Biomechanical Behavior in-vivo for Healthy Eyes Using CorVis ST Tonometry: Stress-Strain Index.
        Front. Bioeng. Biotechnol. 2019; 7: 1-10
        • Chong J.
        • Dupps W.J.
        Corneal biomechanics: Measurement and structural correlations.
        Exp. Eye Res. 2021; 205
        • Akram Abdelazim
        • Joda
        • et al.
        Development and validation of a correction equation for Corvis tonometry.
        Comput. Methods Biomech. Biomed. Engin. 2016; 19: 943-953
        • Hashemi H.
        • et al.
        Long-term Results of an Accelerated Corneal Cross-linking Protocol (18 mW/cm2) for the Treatment of Progressive Keratoconus.
        Am. J. Ophthalmol. 2015; 160: 1164-1170.e1
        • Roberts Cynthia J.
        • et al.
        Introduction of two novel stiffness parameters and interpretation of air Puff-Induced biomechanical deformation parameters with a dynamic scheimpflug analyzer.
        J. Refract. Surg. 2017; 33: 266-273
        • Goldich Y.
        • et al.
        Clinical and corneal biomechanical changes after collagen cross-linking with riboflavin and uv irradiation in patients with progressive keratoconus: Results after 2 years of follow-up.
        Cornea. 2012; 31: 609-614
        • Vinciguerra P.
        • et al.
        Intra- and postoperative variation in ocular response analyzer parameters in keratoconic eyes after corneal cross-linking.
        J. Refract. Surg. 2010; 26: 669-676
        • Elsheikh A.
        • et al.
        Stress free configuration of the human eye.
        Med. Eng. Phys. 2013; 35: 211-216
        • Kirby M.A.
        • et al.
        Optical coherence elastography in ophthalmology.
        J. Biomed. Opt. 2017; 22: 1
        • De Stefano V.S.
        • Ford M.R.
        • Seven I.
        • Dupps W.J.
        Depth-dependent corneal biomechanical properties in normal and keratoconic subjects by optical coherence elastography.
        Transl. Vis. Sci. Technol. 2020; 9: 1-10
        • Wang R.K.
        • Ma Z.
        • Kirkpatrick S.J.
        Tissue Doppler optical coherence elastography for real time strain rate and strain mapping of soft tissue.
        Appl. Phys. Lett. 2006; 89: 1-4
        • Wang R.K.
        • Kirkpatrick S.
        • Hinds M.
        Phase-sensitive optical coherence elastography for mapping tissue microstrains in real time.
        Appl. Phys. Lett. 2007; 90
        • Ambroziński Ł.
        • et al.
        Acoustic micro-tapping for non-contact 4D imaging of tissue elasticity.
        Sci. Rep. 2016; 638967
        • Song S.
        • et al.
        Strategies to improve phase-stability of ultrafast swept source optical coherence tomography for single shot imaging of transient mechanical waves at 16 kHz frame rate.
        Appl. Phys. Lett. 2016; 108: 1-5
        • Kirby M.A.
        • et al.
        Spatial resolution in dynamic optical coherence elastography.
        J. Biomed. Opt. 2019; 24: 1
        • Regnault G.
        • et al.
        Spatial resolution in optical coherence elastography of bounded media.
        Biomed. Opt. Express. 2022; 13: 4851-4869
        • Pitre J.J.
        • et al.
        Nearly-incompressible transverse isotropy (NITI) of cornea elasticity: model and experiments with acoustic micro-tapping OCE.
        Sci. Rep. 2020; 10: 1-14
      1. Petsche, S. J., Chernyak, D., Martiz, J., Levenston, M. E. & Pinsky, P. M. Depth-Dependent Transverse Shear Properties of the Human Corneal Stroma. 53, (2017).

        • Hatami-Marbini H.
        Viscoelastic shear properties of the corneal stroma.
        J. Biomech. 2014; 47: 723-728
        • Hatami-Marbini H.
        • Etebu E.
        Hydration dependent biomechanical properties of the corneal stroma.
        Exp. Eye Res. 2013; 116: 47-54
        • Bekesi N.
        • Dorronsoro C.
        • De La Hoz A.
        • Marcos S.
        Material properties from air puff corneal deformation by numerical simulations on model corneas.
        PLoS One. 2016; 11
        • Boschetti F.
        • Triacca V.
        • Spinelli L.
        • Pandolfi A.
        Mechanical characterization of porcine corneas.
        J. Biomech. Eng. 2012; 134
        • Wollensak G.
        • Spoerl E.
        • Seiler T.
        Stress-strain measurements of human and porcine corneas after riboflavin-ultraviolet-A-induced cross-linking.
        J. Cataract Refract. Surg. 2003; 29: 1780-1785
        • Zeng Y.
        • Yang J.
        • Huang K.
        • Lee Z.
        • Lee X.
        A comparison of biomechanical properties between human and porcine cornea.
        J. Biomech. 2001; 34: 533-537
        • Elsheikh A.
        • Alhasso D.
        Mechanical anisotropy of porcine cornea and correlation with stromal microstructure.
        Exp. Eye Res. 2009; 88: 1084-1091
        • Hatami-Marbini H.
        • Jayaram S.M.
        Effect of UVA/riboflavin collagen crosslinking on biomechanics of artificially swollen corneas.
        Investig. Ophthalmol. Vis. Sci. 2017; 59: 764-770
        • Lanchares E.
        • Del Buey M.A.
        • Cristóbal J.A.
        • Lavilla L.
        • Calvo B.
        Biomechanical property analysis after corneal collagen cross-linking in relation to ultraviolet A irradiation time.
        Graefe’s Arch. Clin. Exp. Ophthalmol. 2011; 249: 1223-1227
        • Aslanides I.M.
        • et al.
        Assessment of UVA-Riboflavin Corneal Cross-Linking Using Small Amplitude Oscillatory Shear Measurements.
        Invest. Ophthalmol. Vis. Sci. 2016; 57: 2240-2245
        • Søndergaard A.P.
        • Ivarsen A.
        • Hjortdal J.
        Corneal resistance to shear force after UVA-riboflavin cross-linking.
        Investig. Ophthalmol. Vis. Sci. 2013; 54: 5059-5069
        • Kirby M.A.
        • et al.
        Delineating corneal elastic anisotropy in a porcine model using non-contact optical coherence elastography and ex vivo mechanical tests.
        Ophthalmol. Sci. 2021; 100058https://doi.org/10.1016/j.xops.2021.100058
        • Baiocchi S.
        • Mazzotta C.
        • Cerretani D.
        • Caporossi T.
        • Caporossi A.
        Corneal crosslinking: Riboflavin concentration in corneal stroma exposed with and without epithelium.
        J. Cataract Refract. Surg. 2009; 35: 893-899
        • Wollensak G.
        • Spoerl E.
        • Seiler T.
        Riboflavin/ultraviolet-A-induced collagen crosslinking for the treatment of keratoconus.
        Am. J. Ophthalmol. 2003; 135: 620-627
        • Petsche S.J.
        • Chernyak D.
        • Martiz J.
        • Levenston M.E.
        • Pinsky P.M.
        Depth-dependent transverse shear properties of the human corneal stroma.
        Investig. Ophthalmol. Vis. Sci. 2012; 53: 873-880
        • Ramirez-Garcia M.A.
        • Sloan S.R.
        • Nidenberg B.
        • Khalifa Y.M.
        • Buckley M.R.
        Depth-Dependent Out-of-Plane Young’s Modulus of the Human Cornea.
        Curr. Eye Res. 2018; 43: 595-604
        • Kirby M.A.
        • et al.
        Probing elastic anisotropy of human skin in vivo with light using non-contact acoustic micro-tapping OCE and polarization sensitive OCT.
        Sci. Rep. 2022; 12: 1-17
        • Tanter M.
        • Touboul D.
        • Gennisson J.L.
        • Bercoff J.
        • Fink M.
        High-resolution quantitative imaging of cornea elasticity using supersonic shear imaging.
        IEEE Trans. Med. Imaging. 2009; 28: 1881-1893
        • Han Z.
        • et al.
        Optical coherence elastography assessment of corneal viscoelasticity with a modified Rayleigh-Lamb wave model.
        J. Mech. Behav. Biomed. Mater. 2017; 66: 87-94
        • Zvietcovich F.
        • et al.
        Longitudinal shear waves for elastic characterization of tissues in optical coherence elastography.
        Biomed. Opt. Express. 2019; 10: 3699
        • Hatami-Marbini H.
        • Rahimi A.
        Stiffening effects of riboflavin/UVA corneal collagen cross-linkingis hydration dependent.
        J. Biomech. 2015; 48: 1052-1057
        • Del Buey M.A.
        • et al.
        Immediate effect of ultraviolet - A collagen cross-linking therapy on the biomechanics and histology of the human cornea.
        J. Refract. Surg. 2015; 31: 70-71
        • Winkler M.
        • et al.
        Nonlinear optical macroscopic assessment of 3-D corneal collagen organization and axial biomechanics.
        Investig. Ophthalmol. Vis. Sci. 2011; 52: 8818-8827
      2. Smolek, M. K. & Klyce, S. D. Is Keratoconus a True Ectasia? 118, (2000).

        • Hayes S.
        • Boote C.
        • Tuft S.J.
        • Quantock A.J.
        • Meek K.M.
        A study of corneal thickness, shape and collagen organisation in keratoconus using videokeratography and X-ray scattering techniques.
        Exp. Eye Res. 2007; 84: 423-434
        • Fan R.
        • Chan T.C.Y.
        • Prakash G.
        • Jhanji V.
        Applications of corneal topography and tomography: a review.
        Clin. Exp. Ophthalmol. 2018; 46: 133-146
        • Hatami-Marbini H.
        • Rahimi A.
        The relation between hydration and mechanical behavior of bovine cornea in tension.
        J. Mech. Behav. Biomed. Mater. 2014; 36: 90-97
        • Seiler T.G.
        • Shao P.
        • Frueh B.E.
        • Yun S.H.
        • Seiler T.
        The influence of hydration on different mechanical moduli of the cornea.
        Graefe’s Arch. Clin. Exp. Ophthalmol. 2018; 256: 1653-1660
        • Kazaili A.
        • Geraghty B.
        • Akhtar R.
        Microscale assessment of corneal viscoelastic properties under physiological pressures.
        J. Mech. Behav. Biomed. Mater. 2019; 100103375
        • Kontadakis G.A.
        • et al.
        In vitro effect of corneal collagen cross-linking on corneal hydration properties and stiffness.
        Graefe’s Arch. Clin. Exp. Ophthalmol. 2013; 251: 543-547
        • Wollensak G.
        • Aurich H.
        • Pham D.T.
        • Wirbelauer C.
        Hydration behavior of porcine cornea crosslinked with riboflavin and ultraviolet A.
        J. Cataract Refract. Surg. 2007; 33: 516-521
        • Hatami-Marbini H.
        • Rahimi
        A. Interrelation of Hydration, Collagen Cross-Linking Treatment, and Biomechanical Properties of the Cornea.
        Curr. Eye Res. 2016; 41: 616-622
        • Elsheikh A.
        • et al.
        Assessment of corneal biomechanical properties and their variation with age.
        Curr. Eye Res. 2007; 32: 11-19
        • Alenezi B.
        • Kazaili A.
        • Akhtar R.
        • Radhakrishnan H.
        Corneal biomechanical properties following corneal cross-linking: Does age have an effect?.
        Exp. Eye Res. 2022; 214
        • Elsheikh A.
        • Alhasso D.
        • Rama P.
        Biomechanical properties of human and porcine corneas.
        Exp. Eye Res. 2008; 86: 783-790
        • Kohlhaas M.
        • et al.
        Biomechanical evidence of the distribution of cross-links in corneastreated with riboflavin and ultraviolet a light.
        J. Cataract Refract. Surg. 2006; 32: 279-283
        • Scott McCall A.
        • et al.
        Mechanisms of corneal tissue cross-linking in response to treatment with topical riboflavin and Long-Wavelength Ultraviolet Radiation (UVA).
        Investig. Ophthalmol. Vis. Sci. 2010; 51: 129-138
        • Beshtawi I.M.
        • O’Donnell C.
        • Radhakrishnan H.
        Biomechanical properties of corneal tissue after ultraviolet-A-riboflavin crosslinking.
        J. Cataract Refract. Surg. 2013; 39: 451-462
        • Hersh P.S.
        • et al.
        United States Multicenter Clinical Trial of Corneal Collagen Crosslinking for Keratoconus Treatment.
        Ophthalmology. 2017; 124: 1259-1270
      3. U.S. Department of Health Education And Welfare. Ocular Ultraviolet Effects from 295nm to 400nm in the Rabbit Eye. NIOSH Research Report (1977).

        • Caruso C.
        • et al.
        Customized Corneal Cross-Linking- A Mathematical Model.
        Cornea. 2017; 36: 600-604
        • Friedman M.D.
        • et al.
        Advanced corneal cross-linking system with fluorescence dosimetry.
        J. Ophthalmol. 2012; 2012
        • Seiler T.G.
        • et al.
        Two-photon fluorescence microscopy for determination of the riboflavin concentration in the anterior corneal stroma when using the dresden protocol.
        Investig. Ophthalmol. Vis. Sci. 2015; 56: 6740-6746
      4. Schumacher, S., Mrochen, M., Wernli, J., Bueeler, M. & Seiler, T. Optimization Model for UV-Riboflavin Corneal Cross-linking. 762–769 (2012). doi:10.1167/iovs.11-8059

        • Spoerl E.
        • Mrochen M.
        • Sliney D.
        • Trokel S.
        • Seiler T.
        Safety of UVA-Riboflavin Cross-Linking of the Cornea.
        Cornea. 2007; 26: 385-389
        • Wollensak G.
        • Spoerl E.
        • Reber F.
        • Seiler T.
        Keratocyte cytotoxicity of riboflavin/UVA-treatment in vitro.
        Eye. 2004; 18: 718-722
        • Wollensak G.
        • Spoerl E.
        • Wilsch M.
        • Seiler T.
        Endothelial cell damage after riboflavin-ultraviolet-A treatment in the rabbit.
        J. Cataract Refract. Surg. 2003; 29: 1786-1790